Super ability in against electrosurgical interference. Neither in cutting nor in burning, ECG waveform and HR can be influenced, very short recovering time after defibrillation, especially suitable for use in emergency or operating room.


Environment
  • Temperature: (+5 ~ +45)℃;
  • Operation & Storage: (-20 ~ +55)℃
  • Relative humidity : ≤ 80 %
  • Operation & Storage:≤ 93 %
  • Atmospheric pressure: (700 ~ 1060)hPa
  • Operation & Storage:(500 ~ 1060)hPa
  • Adapt to power: (100 ~ 240)V a.c. 50/60Hz,
  • Maximum input power: 80VA
  • Fuse: T 1.6 A
Continuous working time:
  • Network power working time : ≥ 8 hours
  • Internal power continuous working: > 1 hour (In premise of full battery, and measuring blood pressure every 10 minutes)
  • Signal Interface:( External monitor): standard VGA port (optional)
  • PRINTER(External printer): PC/AT Parallel Interface
  • NET (External central monitor): RJ45 port
  • LINE(External telephone line): MODEM port

ECG
  • Heart rate(HR)
  • Measurement range: 30BPM ~ 300BPM;
  • Accuracy: 1 BPM or ±1 %.
  • Heart rate alarm settings And tolerance: upper limit (60 ~ 300) bpm; tolerance::± 10% Lower limit (10 ~ 200) bpm; tolerance : ± 10%
  • Noise level: ≤ 30μV
  • ECG frequency response: filtering methods: (1 ~ 25) HzNon-filtered mode: (0.05 ~ 100) Hz The ECG Input circuit current: ≤ 0.1μAECG sensitivity: continuous adjustable
  • The scanning speed of the sub-file and error:Scan rate can be divided into three files for selection: 12.5mm / s, 25mm / s, 50mm / sThe error of scanning rate: ≤ ± 10%
NIBP
Measurement range:
  • Systolic(SBP):(5.3 ~ 36.0)kPa,or(40 ~ 270 )mmHg
  • Diastolic (DBP):(2.7 ~ 26.7)kPa,or(20 ~ 200 )mmHg
  • Mean (MAP):(4.0 ~ 29.3)kPa,or(30 ~ 220 )mmHg
  • Accuracy : ≤ ± 0.4 kPa (3 mmHg) or ±2%,
IBP
  • Measurement range: -1.3~40KPa(-10 ~ 300mmHg)
  • Channel: 2 channelsTransducer sensitivity: 5MV/V/mmHg
  • Unit display: kPa or mmHg( selectable)
Respiration Rate
  • Measurement range: 0 BPM ~ 120 BPM;
  • Accuracy: ±1 BPM or ±5 %
    • Temperature
      • Number of channels: 2
      • Temperature probe configuration: Intracavitary probe: 1 unit; Surface probe: 1 unit
      • Measurement range: 25.0 ℃~ 45.0℃
      • Accuracy: ± 0.2 ℃
      • Response time: ≤ 3 min.
      SpO2
      • Display type: Waveform, data
      • Measurement range: 0~100%
      • Resolution: 1 BPM
      • Pulse rate range: 30~250bpm
      • Accuracy: ±1bpm
      Pulse Rate
      • Measurement range: (30 ~ 250)bpm;
      • Accuracy: 1BPM or ±2%
      • Trend review:Can Review history data of 360 hours
      CO2:
      • Measurement range: 0﹪ ~ 10.0﹪ ( Carbon dioxide inhalation and carbon dioxide exhalation)
      • Accuracy: ±2mmHg(<5﹪ Measured values)
      • 10﹪(>5﹪ Measured values)
      Standard configuration ECG, NIBP, SpO2, HR/PR, 2-RESP, 2-TEMP, rechargeable battery
      Optional configuration Built-in printer, Nellcor SpO2,2-IBP, ETCO2(main stream or side stream), Wireless network, Wall Mount, Moving cart, 12 volts Power supply
Environment
  • Temperature: (+5 ~ +45)℃;
  • Operation & Storage: (-20 ~ +55)℃
  • Relative humidity : ≤ 80 %
  • Operation & Storage:≤ 93 %
  • Atmospheric pressure: (700 ~ 1060)hPa
  • Operation & Storage:(500 ~ 1060)hPa
  • Adapt to power: (100 ~ 240)V a.c. 50/60Hz,
  • Maximum input power: 80VA
  • Fuse: T 1.6 A
Continuous working time:
  • Network power working time : ≥ 8 hours
  • Internal power continuous working: > 1 hour (In premise of full battery, and measuring blood pressure every 10 minutes)
  • Signal Interface:( External monitor): standard VGA port (optional)
  • PRINTER(External printer): PC/AT Parallel Interface
  • NET (External central monitor): RJ45 port
  • LINE(External telephone line): MODEM port,